§84.173. Exhalation valve leakage test.
42 C.F.R. § 84.173
Dry exhalation valves and valve seats will be subjected to a suction of 25 mm water-column height while in a normal operating position.
Leakage between the valve and valve seat must not exceed 30 mL per minute.