§1.1. Creation and authority. — Inbound Citations
40 C.F.R. § 1.1
Statutory Authority
Cited by 44 regulations in release Current.
Citations to 40 C.F.R. § 1.1 as a whole
-
-
(B) Rule R307-110-17, which incorporates revisions to portions of the Utah State Implementation Plan, Section IX, “Control Measures for Area and Point Sources,” Part H, “Emission Limits,” as adopted on June 5, 2002, by the Utah Air Quality, and State effective on September 5, 2002. (Section IX, Part H of the Utah SIP was formerly designated Section 9, Appendix A. The revisions to Section IX, Part H we are incorporating by reference with this action replace the following sections of Section 9, Appendix A of the previously approved Utah SIP: Section 1.1 (General Requirements (Utah County)) and all subsections thereof; Section 1.2 (Particulate Emission Limitations (company specific)) and all subsections thereof.)
-
(A) Air Pollution Control Board, Article 7 SULFUR DIOXIDE RULES, Rule 1.1 Sulfur Dioxide Emission Limitations, sections 326 IAC 7-1.1-1, “Applicability”, 326 IAC 7-1.1-2 “Sulfur Dioxide Emission Limitations”, and 326 IAC 7-2-1 “Reporting Requirements: Methods to Determine Compliance”; newly created 326 IAC 7-4.1, “Lake County Sulfur Dioxide Emission Limitations”, adopted by the Indiana Air Pollution Control Board on March 2, 2005. Filed with the Secretary of State May 25, 2005, effective June 24, 2005.
-
(i) If the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable.
-
(i) If the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable.
-
(i) If the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable.
-
(4) Method 18 of appendix A-6 to this part to determine concentration of TOC in the control device outlet and the concentration of TOC in the inlet when the reduction efficiency of the control device is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 may not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.(i) The sampling time for each run shall be 1 hour in which either an integrated sample or four grab samples shall be taken. If grab sampling is used then the samples shall be taken at 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) shall be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) shall be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and shall be computed for each run using the following equation:
-
(4) Method 18 of appendix A-6 to this part to determine concentration of TOC in the control device outlet or in the outlet of the final recovery device in a recovery system, and to determine the concentration of TOC in the inlet when the reduction efficiency of the control device or recovery system is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.(i) The sampling time for each run must be 1 hour in which either an integrated sample or at least four grab samples must be taken. If grab sampling is used then the samples must be taken at 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) must be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) must be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and must be computed for each run using the following equation:
-
(4) Method 18 of appendix A-6 to this part to determine the concentration of TOC in the control device outlet and the concentration of TOC in the inlet when the reduction efficiency of the control device is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 may not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.(i) The sampling time for each run shall be 1 hour in which either an integrated sample or four grab samples shall be taken. If grab sampling is used then the samples shall be taken at 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) shall be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) shall be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and shall be computed for each run using the following equation:
-
(4) Method 18 of appendix A-6 to this part to determine the concentration of TOC in the control device outlet or in the outlet of the final recovery device in a recovery system, and to determine the concentration of TOC in the inlet when the reduction efficiency of the control device or recovery system is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18, if the target compounds are all known and are all listed in section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 may not be used for methane and ethane; and ASTM D6420-18 must not be used as a total VOC method.(i) The sampling time for each run must be 1 hour in which either an integrated sample or at least four grab samples must be taken. If grab sampling is used then the samples must be taken at 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) must be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) must be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and must be computed for each run using the following equation:
-
(4) Method 18 of appendix A-6 to this part to determine the concentration of TOC in the control device outlet and the concentration of TOC in the inlet when the reduction efficiency of the control device is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18 of appendix A-6 to this part, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 may not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.(i) The minimum sampling time for each run shall be 1 hour in which either an integrated sample or four grab samples shall be taken. If grab sampling is used, then the samples shall be taken at approximately 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) shall be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) shall be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and shall be computed for each run using the following equation:
-
(4) Method 18 of appendix A-6 to this part to determine the concentration of TOC in the control device outlet or in the outlet of the final recovery device in a recovery system, and to determine the concentration of TOC in the inlet when the reduction efficiency of the control device or recovery system is to be determined. ASTM D6420-18 (incorporated by reference, see § 60.17) may be used in lieu of Method 18, if the target compounds are all known and are all listed in section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 may not be used for methane and ethane; and ASTM D6420-18 must not be used as a total VOC method.(i) The minimum sampling time for each run must be 1 hour in which either an integrated sample or at least four grab samples must be taken. If grab sampling is used, then the samples must be taken at approximately 15-minute intervals.(ii) The emission reduction (R) of TOC (minus methane and ethane) must be determined using the following equation:(iii) The mass rates of TOC (Ei, Eo) must be computed using the following equations:(iv) The TOC concentration (CTOC) is the sum of the individual components and must be computed for each run using the following equation:
-
(vi) Wastewater storage tanks are covered under the wastewater provisions.Surface impoundment means a waste management unit which is a natural topographic depression, manmade excavation, or diked area formed primarily of earthen materials (although it may be lined with manmade materials), which is designed to hold an accumulation of liquid wastes or waste containing free liquids. A surface impoundment is used for the purpose of treating, storing, or disposing of wastewater or residuals, and is not an injection well. Examples of surface impoundments are equalization, settling, and aeration pits, ponds, and lagoons.Surge control vessel means feed drums, recycle drums, and intermediate vessels. Surge control vessels are used within a chemical manufacturing process unit when in-process storage, mixing, or management of flow rates or volumes is needed to assist in production of a product.Table 8 compound means a compound listed in table 8 to subpart G of this part.Table 9 compound means a compound listed in table 9 to subpart G of this part.Temperature monitoring device means a unit of equipment used to monitor temperature and having a minimum accuracy of (a) ±1 percent of the temperature being monitored expressed in degrees Celsius (( °C) or (b) ±0.5 degrees ( °C), whichever is greater.The 33/50 program means a voluntary pollution prevention initiative established and administered by the EPA to encourage emissions reductions of 17 chemicals emitted in large volumes by industrial facilities. The EPA Document Number 741-K-92-001 provides more information about the 33/50 program.Total organic compounds (TOC), as used in the process vents provisions, means those compounds measured according to the procedures of Method 18 of appendix A-6 to 40 CFR part 60, ASTM D6420-18 (incorporated by reference, see § 63.14) may be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.Total resource effectiveness index value or TRE index value means a measure of the supplemental total resource requirement per unit reduction of organic HAP associated with a process vent stream, based on vent stream flow rate, emission rate of organic HAP, net heating value, and corrosion properties (whether or not the vent stream contains halogenated compounds), as quantified by the equations given under § 63.115.Transfer operation means the loading, into a tank truck or railcar, of organic liquids that contain one or more of the organic hazardous air pollutants listed in table 2 of this subpart from a transfer rack (as defined in this section). Transfer operations do not include loading at an operating pressure greater than 204.9 kilopascals. For each source as defined in this section, the greater than 204.9 kilopascals exemption in this definition no longer applies on and after July 15, 2027.Transfer rack means the collection of loading arms and loading hoses, at a single loading rack, that are assigned to a chemical manufacturing process unit subject to this subpart according to the procedures specified in § 63.100(h) and are used to fill tank trucks and/or railcars with organic liquids that contain one or more of the organic hazardous air pollutants listed in table 2 to this subpart. Transfer rack includes the associated pumps, meters, shutoff valves, relief valves, and other piping and valves. Transfer rack does not include:
-
(2) Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60 of this chapter, respectively, or Method 320 of appendix A to this part must be used to measure concentration. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18 or Method 320, if the target compounds are all known and are all listed in section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.
-
(3) To determine compliance with the 20 parts per million by volume total organic HAP limit in § 63.113(a)(2) of this subpart, the owner or operator shall use Method 18 of appendix A-6 to part 60 of this chapter to measure either TOC minus methane and ethane or total organic HAP. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. Alternatively, any other method or data that has been validated according to the applicable procedures in Method 301 of appendix A of this part, may be used. The following procedures shall be used to calculate parts per million by volume concentration, corrected to 3 percent oxygen:(i) The minimum sampling time for each run shall be 1 hour in which either an integrated sample or a minimum of four grab samples shall be taken. If grab sampling is used, then the samples shall be taken at approximately equal intervals in time, such as 15 minute intervals during the run.(ii) The concentration of either TOC (minus methane or ethane) or total organic HAP shall be calculated according to paragraph (c)(3)(ii)(A) or (c)(3)(ii)(B) of this section.(A) The TOC concentration (CTOC) is the sum of the concentrations of the individual components and shall be computed for each run using the following equation:(B) The total organic HAP concentration (CHAP) shall be computed according to the equation in paragraph (c)(3)(ii)(A) of this section except that only the organic HAP species shall be summed. The list of organic HAP's is provided in table 2 of subpart F of this part.(iii) The concentration of TOC or total organic HAP shall be corrected to 3 percent oxygen if a combustion device is the control device.(A) Method 3A of appendix A-2 to part 60 of this chapter or the manual method in ANSI/ASME PTC 19.10-1981 (incorporated by reference, see § 63.14) shall be used to determine the oxygen concentration (%O2d). The samples shall be taken during the same time that the TOC (minus methane or ethane) or total organic HAP samples are taken.(B) The concentration corrected to 3 percent oxygen (Cc) shall be computed using the following equation:
-
(E) ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18 of appendix A-6 to part 60 of this chapter, if the target compounds are all known and are all listed in section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.
-
(12) ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18 of appendix A-6 to part 60 of this chapter, if the target compounds are all known and are all listed in section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.
-
(ii) Achieve an outlet total organic compound concentration, less methane and ethane, or total organic hazardous air pollutants concentration of 20 parts per million by volume on a dry basis corrected to 3 percent oxygen. The owner or operator shall use either Method 18 of appendix A-6 to part 60 of this chapter, any other method or data that has been validated according to the applicable procedures in Method 301 of appendix A of this part, or ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method; or
-
(A) Method 18 of appendix A-6 to part 60 of this chapter or Method 320 of appendix A to this part may be used to determine the HAP concentration in any control device efficiency determination. ASTM D6420-18 (Incorporated by reference, see § 63.14) may also be used in lieu of Method 18 or Method 320, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 shall not be used for methane, ethane, or as a total VOC method.
-
(4) The concentration of table 8 and/or table 9 compounds exiting the combustion treatment process in any vented gas stream shall be determined as provided in this paragraph. Samples may be grab samples or composite samples. Samples shall be taken at approximately equally spaced time intervals over a 1-hour period. Each 1-hour period constitutes a run, and the performance test shall consist of a minimum of 3 runs. Concentration measurements shall be determined using Method 18 of appendix A-6 to 40 CFR part 60. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. Alternatively, any other test method validated according to the procedures in Method 301 of appendix A to this part may be used.
-
(2) Determine the concentration of benzene and toluene for each stack using Method 18 in to appendix A-6 to 40 CFR part 60 to determine the concentration of benzene and toluene; or as an alternative ASTM D6420-99 (Reapproved 2010) (incorporated by reference, see § 63.14), may be used provided that the target compound(s) are those listed in section 1.1 of ASTM D6420-99 (Reapproved 2010) as measurable; the target compounds do not include methane and ethane because their atomic mass is less than 35; and the test results are not a total VOC method. Each test must consist of three separate runs. The duration of sampling must be at least two hours per run.
-
(1) Each piece of equipment within a process unit that can reasonably be expected to contain equipment in organic HAP service is presumed to be in organic HAP service unless an owner or operator demonstrates that the piece of equipment is not in organic HAP service. For a piece of equipment to be considered not in organic HAP service, it must be determined that the percent organic HAP content can be reasonably expected not to exceed 5 percent by weight on an annual average basis. For purposes of determining the percent organic HAP content of the process fluid that is contained in or contacts equipment, Method 18 of 40 CFR part 60, appendix A shall be used. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.
-
(ii) You may use ASTM D6420-18 (incorporated by reference, see § 63.14), to determine compliance with the total organic HAP emission limit if the target concentration for each HAP is between 150 parts per billion by volume and 100 ppmv and either of the conditions specified in paragraph (b)(2)(ii)(A) or (B) of this section exists. For target compounds not listed in Section 1.1 of ASTM D6420-18 and not amenable to detection by mass spectrometry, you may not use ASTM D6420-18.(A) The target compounds are those listed in Section 1.1 of ASTM D6420-18 (incorporated by reference, see § 63.14); or(B) For target compounds not listed in Section 1.1 of ASTM D6420-18 (incorporated by reference, see § 63.14), but potentially detected by mass spectrometry, you must demonstrate recovery of the compound and the additional system continuing calibration check after each run, as detailed in ASTM D6420-18, Section 10.5.3, must be followed, met, documented, and submitted with the data report, even if there is no moisture condenser used or the compound is not considered water-soluble.
-
(t) When the provisions of § 63.116(c)(3) and (4) specify that Method 18 of appendix A-6 to 40 CFR part 60 must be used, Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively, may be used for the purposes of this subpart. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. The use of Method 25A must conform with the requirements in paragraphs (t)(1) and (2) of this section.(1) The organic HAP used as the calibration gas for Method 25A of appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of the emissions.(2) [Reserved]
-
(iii) Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively, must be used to determine the concentration of TOC or organic HAP, as appropriate. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. The use of Method 25A must conform with the requirements in paragraphs (b)(5)(iii)(A) and (B) of this section.(A) The organic HAP used as the calibration gas for Method 25A of appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of the emissions.
-
(D) Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively must be used to determine the concentration of organic HAP or TOC, as appropriate. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. Alternatively, any other method or data that has been validated according to the applicable procedures in Method 301 of appendix A to this part may be used. The use of Method 25A must conform with the requirements in paragraphs (c)(1)(i)(D)(1) and (2) of this section.(1) The organic HAP used as the calibration gas for Method 25A appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of the emissions.(2) The use of Method 25A appendix A-7 to 40 CFR part 60 is acceptable if the response from the high-level calibration gas is at least 20 times the standard deviation of the response from the zero calibration gas when the instrument is zeroed on the most sensitive scale.
-
(iii) To determine the inlet and outlet total organic HAP concentrations, the owner or operator must use Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. Alternatively, any other method or data that has been validated according to the applicable procedures in Method 301 of appendix A to this part may be used. The minimum sampling time for each run must be in accordance with paragraph (b)(1) of this section, during which either an integrated sample or grab samples shall be taken. If grab sampling is used, then the samples must be taken at approximately equal intervals during the run, with the time between samples no greater than 15 minutes.
-
(iii) To determine compliance with the carbon disulfide concentration limit in paragraph (a) of this section, the owner or operator shall use Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively to measure carbon disulfide. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. Alternatively, any other method or data that has been validated according to the applicable procedures in Method 301 of appendix A to this part, may be used. The following procedures shall be used to calculate carbon disulfide concentration:(A) The minimum sampling time for each run shall be 1 hour, in which either an integrated sample or a minimum of four grab samples shall be taken. If grab sampling is used, then the samples shall be taken at approximately equal intervals in time, such as 15 minute intervals during the run.(B) The concentration of carbon disulfide shall be calculated using Equation 32.
-
(20) When the provisions of § 63.139(c)(1)(ii), § 63.145(d)(4), or § 63.145(i)(2) specify that Method 18 of appendix A-6 to 40 CFR part 60 must be used, Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively, may be used for the purposes of this subpart. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. The use of Method 25A must conform with the requirements in paragraphs (a)(20)(i) and (ii) of this section.(i) The organic HAP used as the calibration gas for Method 25A of appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of the emissions.(ii) The use of Method 25A of appendix A-7 to 40 CFR part 60 is acceptable if the response from the high-level calibration gas is at least 20 times the standard deviation of the response from the zero calibration gas when the instrument is zeroed on the most sensitive scale.
-
(j) When the provisions of subpart H of this part specify that Method 18 of appendix A-6 to 40 CFR part 60 must be used, either Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively, may be used for the purposes of this subpart. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. The use of Method 25A must conform with the requirements in paragraphs (j)(1) and (2) of this section.(1) The organic HAP used as the calibration gas for Method 25A of appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of emissions.(2) The use of Method 25A of appendix A-7 to 40 CFR part 60 is acceptable if the response from the high-level calibration gas is at least 20 times the standard deviation of the response from the zero calibration gas when the instrument is zeroed on the most sensitive scale.
-
(3) For the purposes of the provisions in this section, whenever Method 18 of appendix A-6 to 40 CFR part 60 is specified within the paragraphs of this section or is specified by reference through provisions outside this section, Method 18 or Method 25A of appendices A-6 and A-7 to 40 CFR part 60, respectively, may be used. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method. The use of Method 25A, must conform with the requirements in paragraphs (a)(3)(i) and (ii) of this section.(i) The organic HAP used as the calibration gas for Method 25Aof appendix A-7 to 40 CFR part 60 must be the single organic HAP representing the largest percent by volume of the emissions.(ii) The use of Method 25A of appendix A-7 to 40 CFR part 60, is acceptable if the response from the high-level calibration gas is at least 20 times the standard deviation of the response from the zero calibration gas when the instrument is zeroed on the most sensitive scale.
-
(iii) Method 25A and/or Methods 18 and 25A of appendices A-6 and A-7 to 40 CFR part 60, as appropriate, must be used to determine the concentration of HAP in the streams. ASTM D6420-18 (incorporated by reference, see § 63.14) may also be used in lieu of Method 18, if the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable; ASTM D6420-18 must not be used for methane and ethane; and ASTM D6420-18 may not be used as a total VOC method.
-
(i) If the target compound(s) is listed in Section 1.1 of ASTM D6420-99 and the target concentration is between 150 parts per billion by volume and 100 parts per million by volume.
-
(i) If the target compound(s) is listed in Section 1.1 of ASTM D6420-99 and the target concentration is between 150 parts per billion by volume and 100 parts per million by volume.
-
(i) The target compound(s) are those listed in section 1.1 of ASTM D6420-99 (Reapproved 2010) as measurable;
-
(A) The target compound(s) are those listed in section 1.1 of ASTM D6420-99 (2004);
-
(A) Except for authorities identified as not being delegated, the regulations incorporated in New Jersey's “Toxic Catastrophe Prevention Act Program,” Title 7, Chapter 31, of the New Jersey Administrative Code: Subchapter 1, “General Provisions” (sections 1.1 to 1.10 except for the definition of “What if Checklist”), Subchapter 2, “Hazard Assessment,” Subchapter 3, “Minimum Requirements for a Program 2 TCPA Risk Management Program,” Subchapter 4, “Minimum Requirements for a Program 3 TCPA Risk Management Program,” Subchapter 5, “Emergency Response,” Subchapter 6, “Extraordinarily Hazardous Substances,” Subchapter 7, “Risk Management Plan and TCPA Submission,” and Subchapter 8, “Other Federal Requirements,” (effective July 20, 1998), pertain to the sources affected by 40 CFR part 68 and have been approved under the procedures in §§ 63.91, 63.93 and 63.95 to be implemented and enforced in place of 40 CFR part 68, Subparts A through H, as may be amended.(1) Authorities not delegated:(i) The New Jersey Department of Environmental Protection is not delegated the Administrator's authority to implement and enforce New Jersey's Toxic Catastrophe Prevention Act Program, Title 7, Chapter 31, of the New Jersey Administrative Code, in lieu of the provisions of 40 CFR part 68 as they apply to the regulation of processes that are covered only because they contain regulated quantities of liquid petroleum gases (LPG) regulated under the New Jersey Liquified Petroleum Gas Act of 1950 (N.J.S.A. 21:1B),(ii) Pursuant to § 63.90(c) the New Jersey Department of Environmental Protection is not delegated the Administrator's authority to add or delete substances from the list of substances established under section 112(r) and set forth in 40 CFR 68.130.
-
(D) To measure the total organic regulated material concentration at the outlet of a control device, use Method 18 of 40 CFR part 60, appendix A, or ASTM D6420-18 (incorporated by reference, see § 63.14). If you have a combustion control device, you must first determine which regulated material compounds are present in the inlet gas stream using process knowledge or the screening procedure described in Method 18. In conducting the performance test, analyze samples collected at the outlet of the combustion control device as specified in Method 18 or ASTM D6420-18 for the regulated material compounds present at the inlet of the control device. The method ASTM D6420-18 may be used only under the conditions specified in paragraphs (e)(2)(iii)(D)(1) through (3) of this section.(1) If the target compounds are all known and are all listed in Section 1.1 of ASTM D6420-18 as measurable.(2) ASTM D6420-18 may not be used for methane and ethane.(3) ASTM D6420-18 may not be used as a total VOC method.
Citations to §1.1(b)(97)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.
Citations to §1.1(b)(98)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.
Citations to §1.1(b)(99)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.
Citations to §1.1(b)(127)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.
Citations to §1.1(b)(128)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.
Citations to §1.1(b)(145)
-
(A) Amendments to Oklahoma Air Pollution Control Regulation (OAPCR) 1.1 “Defining Terms Used in Oklahoma Air Pollution Control Regulations” § 1.1(b)(97), § 1.1(b)(98), § 1.1(b)(99) and § 1.1(b)(145), as adopted October 11, 1989, by the Oklahoma State Board of Health and effective May 25, 1990. Amendments to OAPCR 1.1, § 1.1(b)(127), and § 1.1(b)(128), as adopted March 23, 1989, by the Oklahoma State Board of Health and effective June 11, 1989.