§35.643. Periodic spot-checks for remote afterloader units. — Inbound Citations
10 C.F.R. § 35.643
Statutory Authority
Cited by 7 regulations in release Current.
Citations to 10 C.F.R. § 35.643 as a whole
-
-
-
-
(a) A licensee shall retain a record of each spot-check for remote afterloader units required by § 35.643 for 3 years.
-
(b) In addition to the periodic spot-checks required by § 35.643, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of—(1) Electrical interlocks on treatment area access points;(2) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;(3) Viewing and intercom systems;(4) Applicators, source transfer tubes, and transfer tube-applicator interfaces;(5) Radiation monitors used to indicate room exposures;(6) Source positioning (accuracy); and
-
(b) The approved information collection requirements contained in this part appear in §§ 35.6, 35.12, 35.13, 35.14, 35.19, 35.24, 35.26, 35.27, 35.40, 35.41, 35.50, 35.51, 35.55, 35.60, 35.61, 35.63, 35.67, 35.69, 35.70, 35.75, 35.80, 35.92, 35.190, 35.204, 35.290, 35.310, 35.315, 35.390, 35.392, 35.394, 35.396, 35.404, 35.406, 35.410, 35.415, 35.432, 35.433, 35.490, 35.491, 35.590, 35.604, 35.605, 35.610, 35.615, 35.630, 35.632, 35.633, 35.635, 35.642, 35.643, 35.645, 35.647, 35.652, 35.655, 35.690, 35.1000, 35.2024, 35.2026, 35.2040, 35.2041, 35.2060, 35.2061, 35.2063, 35.2067, 35.2070, 35.2075, 35.2080, 35.2092, 35.2204, 35.2310, 35.2404, 35.2406, 35.2432, 35.2433, 35.2605, 35.2610, 35.2630, 35.2632, 35.2642, 35.2643, 35.2645, 35.2647, 35.2652, 35.2655, 35.3045, 35.3047, 35.3067, and 35.3204.
Citations to §35.643(b)
-
(c) A licensee shall retain a copy of the procedures required by § 35.643(b) until the licensee no longer possesses the remote afterloader unit.