§35.2642. Records of periodic spot-checks for teletherapy units.

10 C.F.R. § 35.2642

(a)
A licensee shall retain a record of each periodic spot-check for teletherapy units required by § 35.642 for 3 years.
(b)
The record must include—
(1)
The date of the spot-check;
(2)
The manufacturer's name, model number, and serial number of the teletherapy unit, source and instrument used to measure the output of the teletherapy unit;
(3)
An assessment of timer linearity and constancy;
(4)
The calculated on-off error;
(5)
A determination of the coincidence of the radiation field and the field indicated by the light beam localizing device;
(6)
The determined accuracy of each distance measuring and localization device;
(7)
The difference between the anticipated output and the measured output;
(8)
Notations indicating the operability of each entrance door electrical interlock, each electrical or mechanical stop, each source exposure indicator light, and the viewing and intercom system and doors; and
(9)
The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
(c)
A licensee shall retain a copy of the procedures required by § 35.642(b) until the licensee no longer possesses the teletherapy unit.