§84.90. Breathing resistance test; inhalation.
42 C.F.R. § 84.90
Resistance to inhalation airflow will be measured in the facepiece or mouthpiece while the apparatus is operated by a breathing machine as described in § 84.88.
The inhalation resistance of open-circuit apparatus shall not exceed 32 mm. (1.25 inch) water-column height (at a flow rate of 120 liters per minute).
The inhalation resistance of closed-circuit apparatus shall not exceed the difference between exhalation resistance (§ 84.91(e)) and 10 cm. (4 inches) water-column height.
Notes, amendments, and revision history
Authority
Authority: 29 U.S.C. 651 et seq.; 30 U.S.C. 3, 5, 7, 811, 842(h), 844.
Source