§84.154. Airflow resistance test; Type B and Type BE supplied-air respirators; minimum requirements.

42 C.F.R. § 84.154

(a)
Airflow resistance shall be determined when the respirator is completely assembled with the respiratory-inlet covering and the hose in the maximum length to be considered for approval, coiled in loops 1.5 to 2.1 m. (5 to 7 feet) in diameter.
(b)
Airflow resistance shall not exceed 38 mm. (1.5 inches) of water-column height to air drawn at the flow rate of 85 liters (3 cubic feet) per minute.
(c)
The exhalation resistance shall not exceed 25 mm. (1 inch) of water-column height at this flow rate.