§1065.905. General provisions. — Inbound Citations
40 C.F.R. § 1065.905
Statutory Authority
Cited by 3 regulations in release Current.
Citations to §1065.905(b)
-
(a) . We recommend that you use PEMS that meet the specifications of subpart C of this part. For unrestricted use of PEMS in a laboratory or similar environment, use a PEMS that meets the same specifications as each lab instrument it replaces. For field testing or for testing with PEMS in a laboratory or similar environment, under the provisions of § 1065.905(b), the specifications in the following table apply instead of the specifications in Table 1 of § 1065.205:
Citations to §1065.905(d)(2)
-
(1) Gravimetric PM measurement must meet the laboratory measurement requirements of this part 1065, noting that there are specific exceptions to some laboratory requirements and specification for field testing given in § 1065.905(d)(2). In addition to those exceptions, field testing does not require you to verify proportional flow control as specified in § 1065.545. Note also that the linearity requirements of § 1065.307 apply only as specified in this section.
Citations to §1065.905(e)
-
(1) Follow the laboratory test procedures specified in this part 1065, according to § 1065.905(e).