§900.12. Quality standards. — Inbound Citations

21 C.F.R. § 900.12

Cited by 49 regulations in release Current.

Citations to 21 C.F.R. § 900.12 as a whole

  • (a) After October 1, 1994, a certificate issued by FDA is required for lawful operation of all mammography facilities subject to the provisions of this subpart. To obtain a certificate from FDA, facilities are required to meet the quality standards in § 900.12 and to be accredited by an approved accreditation body or other entity as designated by FDA.
  • (2) Has failed to comply with the standards of § 900.12;
  • (1) The facility will not be operated in accordance with standards established under § 900.12;
  • (a) Upon application by a qualified party as defined in paragraph (b) of this section, FDA may approve an alternative to a quality standard under § 900.12, when the agency determines that:
    (1) The proposed alternative standard will be at least as effective in assuring quality mammography as the standard it proposes to replace, and
    (i) Is too limited in its applicability to justify an amendment to the standard; or
    (ii) Offers an expected benefit to human health that is so great that the time required for amending the standard would present an unjustifiable risk to the human health; and
    (3) The granting of the alternative is in keeping with the purposes of 42 U.S.C. 263b.
  • (a) State agencies may apply for approval as a certification agency if they have standards at least as stringent as those of § 900.12, qualified personnel, adequate resources to carry out the States as Certifiers' responsibilities, and the authority to enter into a legal agreement with FDA to accept these responsibilities.
  • (c) The scope, timeliness, disposition, and technical accuracy of completed inspections and related enforcement activities shall ensure compliance with facility quality standards required under § 900.12.
  • (ii) Detailed description of the accreditation standards the applicant will require facilities to meet and a discussion substantiating their equivalence to FDA standards required under § 900.12;

Citations to §900.12(a)

Citations to §900.12(a)(1)

Citations to §900.12(a)(1)(i)

  • (iii) Must not employ for provision of the professional component of mammography services a physician or physicians for whom the facility has received written notification by FDA that the physician (or physicians) is (or are) in violation of the certification requirements set forth in section 354 of the PHS Act, as implemented by 21 CFR 900.12(a)(1)(i).

Citations to §900.12(a)(2)

  • (pp) Quality control technologist means an individual meeting the requirements of § 900.12(a)(2) who is responsible for those quality assurance responsibilities not assigned to the lead interpreting physician or to the medical physicist.

Citations to §900.12(a)(3)

Citations to §900.12(a)(4)

Citations to §900.12(b)

Citations to §900.12(c)(1)(iv)(A)

Citations to §900.12(c)(1)(iv)(B)

Citations to §900.12(c)(1)(iv)(C)

Citations to §900.12(c)(1)(iv)(D)

Citations to §900.12(c)(1)(iv)(E)

Citations to §900.12(c)(1)(iv)(F)

Citations to §900.12(c)(1)(iv)(G)

Citations to §900.12(c)(4)(i)

  • (v) Before a facility closes or ceases to provide mammography services, it must make arrangements for access by patients and healthcare providers to their mammographic records. This access may be provided by the permanent transfer of mammographic records to the patient or the patient's healthcare provider or the transfer of the mammographic records to a facility or other entity that will provide access to patients and healthcare providers. Access to the records must be provided by such other facility or entity for the remainder of the time periods specified in paragraph (c)(4)(i) of this section. If a facility ceases to perform mammography but continues to operate as a medical entity, and is able to satisfy the recordkeeping requirements of paragraphs (c)(4)(i) through (iv) of this section, it may choose to continue to retain the medical records rather than transfer them to another facility, unless such a transfer is requested by, or on behalf of, the patient. The facility must notify its accreditation body and certification agency in writing of the arrangements it has made and must make reasonable efforts to notify all affected patients.

Citations to §900.12(c)(4)(ii)

  • (v) Before a facility closes or ceases to provide mammography services, it must make arrangements for access by patients and healthcare providers to their mammographic records. This access may be provided by the permanent transfer of mammographic records to the patient or the patient's healthcare provider or the transfer of the mammographic records to a facility or other entity that will provide access to patients and healthcare providers. Access to the records must be provided by such other facility or entity for the remainder of the time periods specified in paragraph (c)(4)(i) of this section. If a facility ceases to perform mammography but continues to operate as a medical entity, and is able to satisfy the recordkeeping requirements of paragraphs (c)(4)(i) through (iv) of this section, it may choose to continue to retain the medical records rather than transfer them to another facility, unless such a transfer is requested by, or on behalf of, the patient. The facility must notify its accreditation body and certification agency in writing of the arrangements it has made and must make reasonable efforts to notify all affected patients.

Citations to §900.12(c)(4)(iii)

  • (v) Before a facility closes or ceases to provide mammography services, it must make arrangements for access by patients and healthcare providers to their mammographic records. This access may be provided by the permanent transfer of mammographic records to the patient or the patient's healthcare provider or the transfer of the mammographic records to a facility or other entity that will provide access to patients and healthcare providers. Access to the records must be provided by such other facility or entity for the remainder of the time periods specified in paragraph (c)(4)(i) of this section. If a facility ceases to perform mammography but continues to operate as a medical entity, and is able to satisfy the recordkeeping requirements of paragraphs (c)(4)(i) through (iv) of this section, it may choose to continue to retain the medical records rather than transfer them to another facility, unless such a transfer is requested by, or on behalf of, the patient. The facility must notify its accreditation body and certification agency in writing of the arrangements it has made and must make reasonable efforts to notify all affected patients.

Citations to §900.12(c)(4)(iv)

  • (v) Before a facility closes or ceases to provide mammography services, it must make arrangements for access by patients and healthcare providers to their mammographic records. This access may be provided by the permanent transfer of mammographic records to the patient or the patient's healthcare provider or the transfer of the mammographic records to a facility or other entity that will provide access to patients and healthcare providers. Access to the records must be provided by such other facility or entity for the remainder of the time periods specified in paragraph (c)(4)(i) of this section. If a facility ceases to perform mammography but continues to operate as a medical entity, and is able to satisfy the recordkeeping requirements of paragraphs (c)(4)(i) through (iv) of this section, it may choose to continue to retain the medical records rather than transfer them to another facility, unless such a transfer is requested by, or on behalf of, the patient. The facility must notify its accreditation body and certification agency in writing of the arrangements it has made and must make reasonable efforts to notify all affected patients.

Citations to §900.12(d)

  • (i) The facility shall identify a lead interpreting physician who shall have the general responsibility of ensuring that the quality assurance program meets all requirements of paragraphs (d) through (f) of this section. No other individual shall be assigned or shall retain responsibility for quality assurance tasks unless the lead interpreting physician has determined that the individual's qualifications for, and performance of, the assignment are adequate.
  • (x) Lead interpreting physician means the interpreting physician assigned the general responsibility for ensuring that a facility's quality assurance program meets all of the requirements of § 900.12(d) through (f). The administrative title and other supervisory responsibilities of the individual, if any, are left to the discretion of the facility.

Citations to §900.12(d)(f)

  • (i) The facility shall identify a lead interpreting physician who shall have the general responsibility of ensuring that the quality assurance program meets all requirements of paragraphs (d) through (f) of this section. No other individual shall be assigned or shall retain responsibility for quality assurance tasks unless the lead interpreting physician has determined that the individual's qualifications for, and performance of, the assignment are adequate.

Citations to §900.12(e)

  • (b) Regulations published under §§ 1020.30, 1020.31, and 900.12(e) of this chapter that are relevant to equipment performance should also be consulted for a more complete understanding of the equipment performance requirements.
    (1) Radiographic equipment designed for general purpose or special nonmammography procedures shall not be used for mammography. This prohibition includes systems that have been modified or equipped with special attachments for mammography. This requirement supersedes the implied acceptance of such systems in § 1020.31(f)(3) of this chapter.
    (2) All radiographic equipment used for mammography shall be specifically designed for mammography and shall be certified pursuant to § 1010.2 of this chapter as meeting the applicable requirements of §§ 1020.30 and 1020.31 of this chapter in effect at the date of manufacture.
    (i) All devices used in mammography must have met the applicable FDA premarket authorization requirements for medical devices of that type with that intended use.
    (ii) A mammography unit that is converted from one mammographic modality to another is considered a new unit at the facility under this part and must, prior to clinical use, undergo a mammography equipment evaluation demonstrating compliance with applicable requirements. The facility must also follow its accreditation body's procedures for applying for accreditation of that unit.
    (i) The assembly shall be capable of being fixed in any position where it is designed to operate. Once fixed in any such position, it shall not undergo unintended motion.
    (ii) The mechanism ensuring compliance with paragraph (b)(3)(i) of this section shall not fail in the event of power interruption.
    (i) Systems using screen-film image receptors shall provide, at a minimum, for operation with image receptors of 18 × 24 centimeters (cm) and 24 × 30 cm.
    (ii) Systems using screen-film image receptors shall be equipped with moving grids matched to all image receptor sizes provided.
    (iii) Systems used for magnification procedures shall be capable of operation with the grid removed from between the source and image receptor.
    (5) For any mammography system with a light beam that passes through the x-ray beam-limiting device, the light shall provide an average illumination of not less than 160 lux (15 foot candles) at 100 cm or the maximum source-image receptor distance (SID), whichever is less.
    (i) Systems used to perform noninterventional problem solving procedures shall have radiographic magnification capability available for use by the operator.
    (ii) Systems used for magnification procedures shall provide, at a minimum, at least one magnification value within the range of 1.4 to 2.0.
    (i) When more than one focal spot is provided, the system shall indicate, prior to exposure, which focal spot is selected.
    (ii) When more than one target material is provided, the system shall indicate, prior to exposure, the preselected target material.
    (iii) When the target material and/or focal spot is selected by a system algorithm that is based on the exposure or on a test exposure, the system shall display, after the exposure, the target material and/or focal spot actually used during the exposure.
    (8) All mammography systems shall incorporate a compression device.
    (i) Effective October 28, 2002, each system shall provide:
    (A) An initial power-driven compression activated by hands-free controls operable from both sides of the patient; and
    (A) Systems shall be equipped with different sized compression paddles that match the sizes of all full-field image receptors provided for the system. Compression paddles for special purposes, including those smaller than the full size of the image receptor (for “spot compression”) may be provided. Such compression paddles for special purposes are not subject to the requirements of paragraphs (b)(8)(ii)(D) and (b)(8)(ii)(E) of this section.
    (B) Except as provided in paragraph (b)(8)(ii)(C) of this section, the compression paddle shall be flat and parallel to the breast support table and shall not deflect from parallel by more than 1.0 cm at any point on the surface of the compression paddle when compression is applied.
    (C) Equipment intended by the manufacturer's design to not be flat and parallel to the breast support table during compression shall meet the manufacturer's design specifications and maintenance requirements.
    (D) The chest wall edge of the compression paddle shall be straight and parallel to the edge of the image receptor.
    (E) The chest wall edge may be bent upward to allow for patient comfort but shall not appear on the image.
    (i) Manual selection of milliampere seconds (mAs) or at least one of its component parts (milliapere (mA) and/or time) shall be available.
    (ii) The technique factors (peak tube potential in kilovolt (kV) and either tube current in mA and exposure time in seconds or the product of tube current and exposure time in mAs) to be used during an exposure shall be indicated before the exposure begins, except when automatic exposure controls (AEC) are used, in which case the technique factors that are set prior to the exposure shall be indicated.
    (iii) Following AEC mode use, the system shall indicate the actual kilovoltage peak (kVp) and mAs used during the exposure. The mAs may be displayed as mA and time.
    (i) Each screen-film system shall provide an AEC mode that is operable in all combinations of equipment configuration provided, e.g., grid, nongrid; magnification, nonmagnification; and various target-filter combinations.
    (ii) The positioning or selection of the detector shall permit flexibility in the placement of the detector under the target tissue.
    (A) The size and available positions of the detector shall be clearly indicated at the X-ray input surface of the breast compression paddle.
    (B) The selected position of the detector shall be clearly indicated.
    (iii) The system shall provide means for the operator to vary the selected optical density from the normal (zero) setting.
    (11) For facilities using screen-film units, the facility shall use x-ray film for mammography that has been designated by the film manufacturer as appropriate for mammography. For facilities using hardcopy prints of digital images for transfer, retention, or final interpretation purposes, the facility shall use a type of film designated by the film manufacturer as appropriate for these purposes and compatible with the printer being used.
    (12) The facility shall use intensifying screens for mammography that have been designated by the screen manufacturer as appropriate for mammography and shall use film that is matched to the screen's spectral output as specified by the manufacturer.
    (13) For processing mammography films, the facility shall use chemical solutions that are capable of developing the films used by the facility in a manner equivalent to the minimum requirements specified by the film manufacturer.
    (14) The facility shall make special lights for film illumination, i.e., hot-lights, capable of producing light levels greater than that provided by the view box, available to the interpreting physicians.
    (15) Facilities shall ensure that film masking devices that can limit the illuminated area to a region equal to or smaller than the exposed portion of the film are available to all interpreting physicians interpreting for the facility.
    (16) Systems with image receptor modalities other than screen-film shall demonstrate compliance with quality standards by successful results of quality assurance testing as specified under paragraph (e)(6) of this section.
  • (3) Manufacturers and assemblers of equipment used for mammography may apply for alternatives to the standards of § 900.12(b) and (e).
  • (bb) Mammography equipment evaluation means an onsite assessment of mammography unit or image processor performance by a medical physicist for the purpose of making a preliminary determination as to whether the equipment meets all of the applicable standards in § 900.12(b) and (e).
  • (i) With its initial accreditation application, a mammography equipment evaluation that was performed by a medical physicist no earlier than 6 months before the date of application for accreditation by the facility. Such evaluation shall demonstrate compliance of the facility's equipment with the requirements in § 900.12(e).

Citations to §900.12(e)(1)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(2)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(3)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(4)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(5)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(6)

  • (7) The facility shall verify that mammography units used to produce mammograms at more than one location meet the requirements in paragraphs (e)(1) through (e)(6) of this section. In addition, at each examination location, before any examinations are conducted, the facility shall verify satisfactory performance of such units using a test method that establishes the adequacy of the image quality produced by the unit.

Citations to §900.12(e)(7)

  • (i) After completion of the tests specified in paragraphs (e)(1) through (e)(7) of this section, the facility shall compare the test results to the corresponding specified action limits; or, for nonscreen-film modalities, to the manufacturer's recommended action limits; or, for post-move, preexamination testing of mobile units, to the limits established in the test method used by the facility.

Citations to §900.12(f)

  • (x) Lead interpreting physician means the interpreting physician assigned the general responsibility for ensuring that a facility's quality assurance program meets all of the requirements of § 900.12(d) through (f). The administrative title and other supervisory responsibilities of the individual, if any, are left to the discretion of the facility.

Citations to §900.12(f)(1)

Citations to §900.12(f)(2)

Citations to §900.12(f)(3)

Citations to §900.12(j)

  • (3) Has failed to comply with reasonable requests of FDA, the State certification agency, or the accreditation body for records, information, reports, or materials, including clinical images for an additional mammography review under § 900.12(j), that FDA or the State certification agency believes are necessary to determine the continued eligibility of the facility for a certificate or continued compliance with the standards of § 900.12;
  • (A) Procedures for performing accreditation and reaccreditation clinical image review in accordance with § 900.4(c), random clinical image reviews in accordance with § 900.4(f), and additional mammography review in accordance with § 900.12(j);