US Codex
C.F.R.
Browse by date
Notes

§870.2790. Photoplethysmograph analysis software for over-the-counter use.

21 C.F.R. § 870.2790

(a)
Identification. A photoplethysmograph analysis software device for over-the-counter use analyzes photoplethysmograph data and provides information for identifying irregular heart rhythms. This device is not intended to provide a diagnosis.
(b)
Classification. Class II (special controls). The special controls for this device are:
(1)
Clinical performance testing must demonstrate the performance characteristics of the detection algorithm under anticipated conditions of use.
(2)
Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the detection algorithm and its inputs and outputs.
(3)
Non-clinical performance testing must demonstrate the ability of the device to detect adequate photoplethysmograph signal quality.
(4)
Human factors and usability testing must demonstrate the following:
(i)
The user can correctly use the device based solely on reading the device labeling; and
(ii)
The user can correctly interpret the device output and understand when to seek medical care.
(5)
Labeling must include:
(i)
Hardware platform and operating system requirements;
(ii)
Situations in which the device may not operate at an expected performance level;
(iii)
A summary of the clinical performance testing conducted with the device;
(iv)
A description of what the device measures and outputs to the user; and
(v)
Guidance on interpretation of any results.
Notes, amendments, and revision history

Amendments

[87 FR 6419, Feb. 4, 2022]

Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360 l, 371.

Source

Source: 45 FR 7907, Feb. 5, 1980, unless otherwise noted.

Amendments

[87 FR 6419, Feb. 4, 2022]