21 C.F.R. § 812.20
(b)
Contents. An IDE application shall include, in the following order:
(1)
The name and address of the sponsor.
(2)
A complete report of prior investigations of the device and an accurate summary of those sections of the investigational plan described in
§ 812.25(a) through (e) or, in lieu of the summary, the complete plan. The sponsor shall submit to FDA a complete investigational plan and a complete report of prior investigations of the device if no IRB has reviewed them, if FDA has found an IRB's review inadequate, or if FDA requests them.
(3)
A description of the methods, facilities, and controls used for the manufacture, processing, packing, storage, and, where appropriate, installation of the device, in sufficient detail so that a person generally familiar with good manufacturing practices can make a knowledgeable judgment about the quality control used in the manufacture of the device.
(4)
An example of the agreements to be entered into by all investigators to comply with investigator obligations under this part, and a list of the names and addresses of all investigators who have signed the agreement.
(5)
A certification that all investigators who will participate in the investigation have signed the agreement, that the list of investigators includes all the investigators participating in the investigation, and that no investigators will be added to the investigation until they have signed the agreement.
(6)
A list of the name, address, and chairperson of each IRB that has been or will be asked to review the investigation and a certification of the action concerning the investigation taken by each such IRB.
(7)
The name and address of any institution at which a part of the investigation may be conducted that has not been identified in accordance with
paragraph (b)(6) of this section.
(8)
If the device is to be sold, the amount to be charged and an explanation of why sale does not constitute commercialization of the device.
(10)
Copies of all labeling for the device.
(11)
Copies of all forms and informational materials to be provided to subjects to obtain informed consent.
(12)
Any other relevant information FDA requests for review of the application.
Notes, amendments, and revision history
Amendments
[45 FR 3751, Jan. 18, 1980, as amended at 46 FR 8956, Jan. 27, 1981; 50 FR 16669, Apr. 26, 1985; 53 FR 11252, Apr. 6, 1988; 61 FR 51530, Oct. 2, 1996; 62 FR 40600, July 29, 1997; 64 FR 10942, Mar. 8, 1999; 73 FR 49942, Aug. 25, 2008; 84 FR 68339, Dec. 16, 2019]
Authority
Authority: 21 U.S.C. 331, 351, 352, 353, 355, 360, 360c-360f, 360h-360j, 360hh-360pp, 360rr-360ss, 360bbb-8b, 371, 372, 374, 379e, 381, 382; 42 U.S.C. 216, 241, 262.
Source
Source: 45 FR 3751, Jan. 18, 1980, unless otherwise noted.
Amendments
[45 FR 3751, Jan. 18, 1980, as amended at 46 FR 8956, Jan. 27, 1981; 50 FR 16669, Apr. 26, 1985; 53 FR 11252, Apr. 6, 1988; 61 FR 51530, Oct. 2, 1996; 62 FR 40600, July 29, 1997; 64 FR 10942, Mar. 8, 1999; 73 FR 49942, Aug. 25, 2008; 84 FR 68339, Dec. 16, 2019]