# §84.148. Type C supplied-air respirator, continuous flow class; minimum requirements.

- (a) Respirators tested under this section shall be approved only when they supply respirable air at the pressures and quantities required.
- (b) The pressure at the inlet of the hose connection shall not exceed 863 kN/m. 2 (125 pounds per square inch gage).
- (c) Where the pressure at any point in the supply system exceeds 863 kN/m. 2 (125 pounds per square inch gage), the respirator shall be equipped with a pressure-release mechanism that will prevent the pressure at the hose connection from exceeding 863 kN/m. 2 (125 pounds per square inch gage) under any conditions.