# §84.103. Man tests; performance requirements.

- (a) The apparatus shall satisfy the respiratory requirements of the wearer for the classified service time.
- (b) Fogging of the eyepiece shall not obscure the wearer's vision, and the wearer shall not experience undue discomfort because of fit or other characteristics of the apparatus.
- (c) **When the ambient temperature during testing is 24 ±6 °C.** (75 ±10 °F.), the maximum temperature of inspired air recorded during man tests shall not exceed the following, after correction for deviation from 24 °C. (75 °F.):