# §1065.265. Nonmethane cutter.

- (a) **Application.** You may use a nonmethane cutter to measure CH4 with a FID analyzer. A nonmethane cutter oxidizes all nonmethane hydrocarbons to CO2 and H2O. You may use a nonmethane cutter for raw or diluted exhaust for batch or continuous sampling.
- (b) **System performance.** Determine nonmethane-cutter performance as described in [§ 1065.365](/cfr/40/1065.365.md) and use the results to calculate CH4 or NMHC emissions in [§ 1065.660](/cfr/40/1065.660.md).
- (c) **Configuration.** Configure the nonmethane cutter with a bypass line if it is needed for the verification described in [§ 1065.365](/cfr/40/1065.365.md).
- (d) **Optimization.** You may optimize a nonmethane cutter to maximize the penetration of CH4 and the oxidation of all other hydrocarbons. You may humidify a sample and you may dilute a sample with purified air or oxygen (O2) upstream of the nonmethane cutter to optimize its performance. You must account for any sample humidification and dilution in emission calculations.