# §870.3605. Pacing system analyzer.

- (a) **Identification.** A pacing system analyzer (PSA) is a prescription device that combines the functionality of a pacemaker electrode function tester ([§ 870.3720](/cfr/21/870.3720.md)) and an external pacemaker pulse generator (EPPG) ([§ 870.3600](/cfr/21/870.3600.md)). It is connected to a pacemaker lead and uses a power supply and electronic circuits to supply an accurately calibrated, variable pacing pulse for measuring the patient's pacing threshold and intracardiac R-wave potential. A PSA may be a single, dual, or triple chamber system and can simultaneously deliver pacing therapy while testing one or more implanted pacing leads.
- (b) **Classification.** Class II (special controls). The special controls for this device are:
  - (1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment.
  - (2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source (i.e., battery power, AC mains connections, or both).
  - (3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
    - (i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling;
    - (ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use;
    - (iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices (e.g. data storage, printing), and indicator(s) functionality under intended use conditions; and
    - (iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated.
  - (4) Appropriate software verification, validation, and hazard analysis must be performed.
  - (5) Labeling must include the following:
    - (i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use;
    - (ii) **Connector terminals should be clearly, unambiguously marked on the outside of the PSA.** The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals. Triple chamber devices should clearly identify atrial, right ventricular, and left ventricular terminals;
    - (iii) The labeling must list all pacing modes available in the device;
    - (iv) Labeling must include a detailed description of any special capabilities (e.g., overdrive pacing or automatic mode switching);
    - (v) Labeling must limit the use of external pacing to the implant procedure; and
    - (vi) Appropriate electromagnetic compatibility information must be included.