# §870.1255. Balloon aortic valvuloplasty catheter.

- (a) **Identification.** A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.
- (b) **Classification.** Class II (special controls). The special controls for this device are:
  - (1) The device must be demonstrated to be biocompatible.
  - (2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
  - (3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
  - (4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
  - (5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.