# §35.632. Full calibration measurements on teletherapy units.

- (a) A licensee authorized to use a teletherapy unit for medical use shall perform full calibration measurements on each teletherapy unit—
  - (1) Before the first medical use of the unit; and
  - (2) **Before medical use under the following conditions—**
    - (i) Whenever spot-check measurements indicate that the output differs by more than 5 percent from the output obtained at the last full calibration corrected mathematically for radioactive decay;
    - (ii) Following replacement of the source or following reinstallation of the teletherapy unit in a new location;
    - (iii) Following any repair of the teletherapy unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and
  - (3) **At intervals not exceeding 1 year.**
- (b) To satisfy the requirement of [paragraph (a)](#a) of this section, full calibration measurements must include determination of—
  - (1) The output within ±3 percent for the range of field sizes and for the distance or range of distances used for medical use;
  - (2) The coincidence of the radiation field and the field indicated by the light beam localizing device;
  - (3) The uniformity of the radiation field and its dependence on the orientation of the useful beam;
  - (4) Timer accuracy and linearity over the range of use;
  - (5) On-off error; and
  - (6) **The accuracy of all distance measuring and localization devices in medical use.**
- (c) A licensee shall use the dosimetry system described in [§ 35.630(a)](/cfr/10/35.630.md?p=a) to measure the output for one set of exposure conditions. The remaining radiation measurements required in [paragraph (b)(1)](#b-1) of this section may be made using a dosimetry system that indicates relative dose rates.
- (d) A licensee shall make full calibration measurements required by [paragraph (a)](#a) of this section in accordance with published protocols accepted by nationally recognized bodies.
- (e) A licensee shall mathematically correct the outputs determined in [paragraph (b)(1)](#b-1) of this section for physical decay for intervals not exceeding 1 month for cobalt-60, 6 months for cesium-137, or at intervals consistent with 1 percent decay for all other nuclides.
- (f) Full calibration measurements required by [paragraph (a)](#a) of this section and physical decay corrections required by [paragraph (e)](#e) of this section must be performed by the authorized medical physicist.
- (g) A licensee shall retain a record of each calibration in accordance with [§ 35.2632](/cfr/10/35.2632.md).