# §35.61. Calibration of survey instruments.

- (a) A licensee shall calibrate the survey instruments used to show compliance with this part and [10 CFR Part 20](/cfr/10/part20.md) before first use, annually, and following a repair that affects the calibration. A licensee shall—
  - (1) Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;
  - (2) Calibrate two separated readings on each scale or decade that will be used to show compliance; and
  - (3) **Conspicuously note on the instrument the date of calibration.**
- (b) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
- (c) A licensee shall retain a record of each survey instrument calibration in accordance with [§ 35.2061](/cfr/10/35.2061.md).