# §34.25. Radiation survey instruments.

- (a) The licensee shall keep sufficient calibrated and operable radiation survey instruments at each location where radioactive material is present to make the radiation surveys required by this part and by [10 CFR part 20](/cfr/10/part20.md) of this chapter. Instrumentation required by this section must be capable of measuring a range from 0.02 millisieverts (2 millirems) per hour through 0.01 sievert (1 rem) per hour.
- (b) The licensee shall have each radiation survey instrument required under [paragraph (a)](#a) of this section calibrated—
  - (1) At intervals not to exceed 6 months and after instrument servicing, except for battery changes;
  - (2) For linear scale instruments, at two points located approximately one-third and two-thirds of full-scale on each scale; for logarithmic scale instruments, at mid-range of each decade, and at two points of at least one decade; and for digital instruments, at 3 points between 0.02 and 10 millisieverts (2 and 1000 millirems) per hour; and
  - (3) So that an accuracy within plus or minus 20 percent of the calibration source can be demonstrated at each point checked.
- (c) The licensee shall maintain records of the results of the instrument calibrations in accordance with [§ 34.65](/cfr/10/34.65.md).